System Diagnostics

Select a code below to view quick details, or access the full resolution guide.

Equalizing Pressure System Alert

Equalizing Pressure

Pressure stabilization when the device is idle

System Analysis

Description

Pressure stabilization when the device is idle — equalization to atmospheric pressure.

Possible Causes:

A sterilization pouch may have fallen out of the tray and become lodged at the back of the chamber, blocking the drain outlet
A malfunctioning control board, which misreads pressure levels and falsely triggers this message.

1. Inspect the chamber

Inspect the chamber to ensure no pouch has slipped behind the tray and blocked the outlet and remove it if found.

2. Restart the device

Restart the device and observe behavior.