System Diagnostics

Select a code below to view quick details, or access the full resolution guide.

Code 2 System Alert

Code 2

Steam Generator Overtemperature

Support Level
Tier 1 (User Resolvable)
System Analysis

Description

The temperature in the steam generator has exceeded the operational limit.

This may be due to placing too many instruments inside the chamber — exceeding the recommended load can lead to overheating.

Another possible cause is a malfunction in the steam generator or incorrect interpretation of sensor data by the mainboard. In most cases, this is caused by using improper water.

Follow the resolution protocol below.

1. Check the load weight

Ensure that the sterilized load does not exceed the recommended weight: max. 0.5 kg (S, Beauty) or 0.8 kg (PRO). Repeat the cycle with a lighter load